AlignmentBasedQuantification

Tool Description

QuantBasedAlignment predicts where a peptide ion identified in another run should elute in this run.

Alignment

How does AlignmentBasedQuantification work?

  1. Extract transferred ion chromatograms

    AlignmentBasedQuantification extracts the ion chromatogram (XIC) of every transferred ion around that prediction, decides which of the peaks in the XIC belongs to the ion, and fits that peak as PSMBasedQuantification does.

  2. Learn peak selection per run

    The decision is made by a model that each run learns from the ions it identified itself. The output is a .quant file with the run's own quantifications followed by the transferred ones, which AlignmentBasedQuantStatistics then scores.

  3. Detect peaks in the chromatogram

    The XIC is extracted over ±ScanTimeWindow around PredictedScanTime, baseline corrected, and its peaks are detected with the wavelet or second derivative method of the PSM based quantification.

  4. Define the candidate window

    Candidates are the peaks whose position lies within ±max(MinScanTimeWindow, ScanTimeWindowSdFactor · PredictedScanTimeSd) of the prediction, at most ±ScanTimeWindow. With the defaults this is ±1 min where the alignment is precise and up to ±2 min where it is not.

  5. Describe each candidate

    Each candidate gets the features below.

  6. Select and fit the best peak

    A logistic regression gives each candidate a probability of being the ion's peak. The candidate with the highest probability is fitted.

A peak is always chosen when any candidate lies in the window, so this step never rejects an ion. Ions whose analyte is absent from the run are handled by AlignmentBasedQuantStatistics, which uses the decoy transfers described below.

Feature Description
Scan time distance |candidate scan time − PredictedScanTime| / max(PredictedScanTimeSd, 0.03 min), and its square
Shape correlation correlation of the candidate with the source trace, shifted onto the candidate
Relative height log10 of the candidate height relative to the highest candidate
Width ratio |log2(candidate FWHM / source FWHM)|, using the full widths at half maximum
Number of candidates log2 of the number of candidates in the window
Isotope evidence correlation of the M0 and M+1 traces over the candidate, and log2 (M+1 / M0)
Mass error |median M0 mass error in ppm| around the candidate apex
Ion mobility distance |candidate ion mobility − PredictedIonMobility| / max(PredictedIonMobilitySd, 0.001) (timsTOF only, with ion mobility in 1/K0)

The model is trained per run on the ions the run identified itself and that QuantBasedAlignment also transferred (overlap ions). For these the right answer is known: the candidate within one standard deviation of the apex of the run's own PSM based peak. Only overlap ions with a reliable own fit (|search − apex of the fitted peak| < 2 standard deviations of the fitted peak) are used.

The training is cross-fitted: the peptides are split into 5 folds, and an overlap ion is decided by the model trained on the other four folds. New ions are decided by the model trained on all overlap ions.

A run with fewer than 200 training ions uses a fixed score instead: −0.5 · min(d², 50) + 4 · shape correlation + 2 · log10 relative height, with d the scan time distance.

AlignmentBasedQuant
In the figure four peaks lie in the candidate window. The highest is 0.6 min from the prediction. The model chooses the peak next to the prediction, whose shape matches the source trace. It is the peak of the run's own identification.

An ion that cannot be quantified gets no row, for example when no candidate lies in the window, the XIC is too short, or the peak detection or the fit fails. The reason is recorded in the Status column of <run>.alignfeatures.

m/z window

The tool measures the offset between observed and theoretical m/z in the MS1 spectra of ions the run quantified itself. After removing outliers it fits the offset in ppm as the sum of a function of scan time and a function of m/z. Every trace of an ion (monoisotopic, isotopes, labeled partner) is centred on its theoretical m/z plus the correction at the PredictedScanTime and that m/z.

With MzWindow_Da set to Estimate the half width is 4 standard deviations of the offsets that remain after the correction, the same at every m/z. Fixed sets the half width in Da and keeps the correction.

With fewer than 50 measured ions left, the correction is fitted on the precursor m/z of the .quant file instead, and Estimate uses a half width of 0.05 Da.

Ion mobility

Ion mobility is used when the spectra and the .align file have it. An XIC point is then the sum of the intensity of all peaks inside the m/z window and inside the ion mobility window, as in PSMBasedQuantificationTIMs. Both widths are part of the quantity, so transferred and PSM based quantities are comparable when both tools use the same widths.

The traces of an ion, of its isotopes and of its labeled partner share one window, set by XicExtraction.IonMobilityWindow. The expected ion mobility of the ion is PredictedIonMobility, or its ion mobility in the source run when the .align row has no prediction.

  • Fixed v: ±v around the ion mobility centre of the ion. The centre is the intensity weighted ion mobility of the signal within v of the expected ion mobility, around PredictedScanTime. Without such signal it is the expected ion mobility.
  • Estimate: ±max(0.05, 4 · PredictedIonMobilitySd) 1/K0 around the expected ion mobility, ±0.05 without prediction.

In a transferred row, IonMobility is the intensity weighted ion mobility of the chosen peak, measured within one fitted standard deviation of its apex. IonMobilityCentre is the ion mobility the traces were summed around: the centre with Fixed, the expected ion mobility with Estimate, NaN without ion mobility.

Labeled quantification

With PerformLabeledQuantification other than Unlabeled the partner of the other label is extracted with the same windows. Its peak is the peak that contains the fitted scan time of the ion. With RequirePeakAtScanTime set to false, the peak with the closest apex is taken when none contains it. A partner that cannot be quantified keeps its m/z and traces but gets no quantity, and the row is kept. The ion's own peak is always the candidate chosen by the selection model.

An ion that the run did not quantify itself is not transferred when the PSM based row of its labeled partner has a fitted peak and a quantity of the ion's label. Its Status is labeled partner quantified by its PSM based row.

InputsFiles and command-line flags
Flag Description Comes from
-i .mzlite of the run, or a directory MzMLToMzLite or MzMLToMzLiteIonMobility
-ii .align of the run, or a directory QuantBasedAlignment
-iv .quant of the same run, or a directory PSMBasedQuantification or PSMBasedQuantificationTIMs
-d peptide database (SQLite) PeptideDB
-o output directory, created when missing -
-p the parameter file (JSON) AlignmentBasedQuantificationParams.json
-mf switch: pair the files by base name instead of by position -
-dc switch: write <run>_mzCorrection.html with the m/z offsets and their correction -
-c number of runs processed in parallel, one core each, default 1 -
  • Without -mf the files are paired by position, the files of a directory sorted by name. If the numbers of files differ, the tool stops.
OutputsGenerated data and log files
  • <run>.quant has the layout of the PSM based .quant files. It holds the PSM based rows with a fitted peak, then one row per quantified transferred ion with QuantificationSource = Alignment.
  • These rows pass no quality filter. In these rows Difference_SearchRT_FittedRT is PredictedScanTime minus the fitted scan time. AlignmentScore and AlignmentQValue hold NaN until AlignmentBasedQuantStatistics fills them.
  • <run>.alignfeatures is a tab separated table with one row per .align row. AlignmentBasedQuantStatistics reads it from the directory of the aligned .quant and uses its features. The main columns:
Column Description
StringSequence, GlobalMod, Charge, PepSequenceID, ModSequenceID Peptide Ion
CandidateProbability probability of the chosen candidate, nan with the fixed score
CandidateMargin lead of the chosen candidate over the second best
NCandidates number of candidates
RtZ, ImZ, ShapeCorr, LogRelHeight, AbsLog2FwhmRatio, IsoCorr, IsoRatioLog2, AbsPpmMono the features of the chosen candidate
SelectionModel Logistic or FixedScore
CandidateScanTime position of the chosen candidate, in minutes
ScanTimeWindow, MzWindow, IonMobilityWindow half widths of the windows used
Status quantified, or why the peptide ion was not quantified
  • With DecoyFraction above 0, <run>.aligndecoy holds the decoy transfers: a share of the new ions quantified a second time with PredictedScanTime shifted by 4 to 6 minutes, where their own peak is not.
  • A decoy is skipped when the ion's own peak could lie in its XIC. Each row has the .quant columns, the shifted PredictedScanTime and the .alignfeatures feature columns.
  • AlignmentBasedQuantStatistics uses them to estimate how often a peptide ion whose analyte is absent from the run passes the Q-value cutoff. They never enter <run>.quant.
  • Logs go to AlignmentBasedQuantification_log.txt and one <run>_log.txt per run, with the counts of ions per reason at Info level and each peptide ion that was not quantified at Debug level.

The tool overwrites its .quant and .alignfeatures outputs for the run. If decoy transfers are disabled, it removes an existing .aligndecoy for that run.

ParametersDefault values and tool settings
Parameter Default Description
PerformLabeledQuantification Labeling.N15Labeling Also quantify the partner of the other label. Unlabeled quantifies only the ion itself, for label-free data. N15Labeling, N15LabelingOnly and Labelshift behave the same here.
XicExtraction see below How the XIC is extracted and processed.
BaseLineCorrection Some { MaxIterations = 10; Lambda = 6; P = 0.05 } Asymmetric least squares baseline of the XIC. None switches it off.
MinScanTimeWindow 1.0 Smallest half width of the candidate window, in minutes.
ScanTimeWindowSdFactor 8.0 Half width of the candidate window in units of PredictedScanTimeSd, if larger than MinScanTimeWindow.
DecoyFraction 0.2 Share of the new ions that also get a decoy transfer. 0 writes no decoys.
Seed 1 Seed of the cross-fitting folds and of the choice of the decoys.
RequirePeakAtScanTime true Whether the labeled partner needs a peak that contains the fitted scan time of the ion (see Labeled quantification).

XIC Extraction

XicExtraction field Default Description
ScanTimeWindow 2.0 Half width of the XIC in minutes, and the largest half width of the candidate window.
MzWindow_Da Window.Estimate Estimate measures the m/z window from the run. Window.Fixed 0.01 sets ±0.01 Da.
XicProcessing XicProcessing.Wavelet Peak detection. The wavelet fields are described on the PSMBasedQuantification page. The default BorderPadMethod is Zero. Random is accepted with a warning, because it makes the detected peaks differ between reruns.
TopKPSMs None Not used
IonMobilityWindow Window.Estimate Estimate or Window.Fixed v, in 1/K0 (see Ion mobility).

For timsTOF runs, AlignmentBasedQuantificationTIMsParams.json sets the m/z half width to Window.Fixed 0.008 Da and the ion mobility half width to Window.Fixed 0.025 in 1/K0. These match the defaults in PSMBasedQuantificationTIMs. Suitable windows depend on the instrument and acquisition method.

Parameter generationJSON and parameter-generation examples

From the ProteomIQon repository root, build the library with dotnet build src/ProteomIQon/ProteomIQon.fsproj -c Release. Save the following as generate-params.fsx in that same directory and run dotnet fsi generate-params.fsx. The script writes the JSON file to the system temporary directory.

#r "nuget: FSharp.Stats, 0.6.0"
#r "nuget: Newtonsoft.Json, 13.0.4"
#r "./src/ProteomIQon/bin/Release/net10.0/ProteomIQon.dll"

open ProteomIQon
open ProteomIQon.Domain
open FSharp.Stats.Signal

let waveletParams : WaveletParameters =
    {
        Borderpadding          = None
        BorderPadMethod        = Padding.BorderPaddingMethod.Zero
        InternalPaddingMethod  = Padding.InternalPaddingMethod.LinearInterpolation
        HugeGapPaddingMethod   = Padding.HugeGapPaddingMethod.Zero
        HugeGapPaddingDistance = 100.
        MinPeakDistance        = None
        MinPeakLength          = Some 0.1
        MaxPeakLength          = 1.5
        NoiseQuantile          = 0.01
        MinSNR                 = 0.01
    }

let alignmentBasedQuantificationParams : Dto.AlignmentBasedQuantificationParams =
    {
        // Labeling.Unlabeled for label-free data
        PerformLabeledQuantification = Labeling.N15Labeling
        XicExtraction =
            {
                ScanTimeWindow = 2.
                MzWindow_Da    = Window.Estimate
                XicProcessing  = XicProcessing.Wavelet waveletParams
                TopKPSMs       = None
                IonMobilityWindow = Window.Estimate
            }
        BaseLineCorrection     = Some { MaxIterations = 10; Lambda = 6; P = 0.05 }
        MinScanTimeWindow      = 1.
        ScanTimeWindowSdFactor = 8.
        DecoyFraction          = 0.2
        Seed                   = 1
        RequirePeakAtScanTime  = true
    }

// Replace the temp folder with your project folder.
let outputPath =
    System.IO.Path.Combine(System.IO.Path.GetTempPath(), "AlignmentBasedQuantificationParams.json")

Json.serializeAndWrite outputPath alignmentBasedQuantificationParams

Running the tool

  1. Installing

    Install the tool globally with the .NET CLI.

    dotnet tool install --global ProteomIQon.AlignmentBasedQuantification
  2. Quantify one run

    proteomiqon-alignmentbasedquantification -i path/to/run.mzlite -ii path/to/run.align -iv path/to/run.quant -d path/to/peptides.db -o path/to/output -p path/to/AlignmentBasedQuantificationParams.json
  3. Process every run in a set of directories, paired by file name, four at a time

    proteomiqon-alignmentbasedquantification -i path/to/mzlite -ii path/to/align -iv path/to/quant -d path/to/peptides.db -o path/to/output -p path/to/AlignmentBasedQuantificationParams.json -mf -c 4
  4. All Flags

    proteomiqon-alignmentbasedquantification --help

Limitations

  • MinScanTimeWindow (1 min) and ScanTimeWindow (2 min) are fixed in minutes and do not scale with the gradient length.
  • The fixed score uses the same weights in every run.
  • The right peak must be among the detected peaks. Shoulders that the peak detection merges with a neighbour cannot be chosen separately.