ProteomIQon
Menu

PSMBasedQuantification

Tool Description

PSMBasedQuantification turns the identifications from PSMStatistics into peptide ion intensities.

How does PSMBasedQuantification work?

  1. XIC Extraction

    For every identified peptide ion (sequence, modification state and charge) it takes the scan times of its MS/MS identifications, extracts an ion chromatogram of the monoisotopic m/z in a window around them, detects the peaks in that trace and fits the peak that contains the weighted scan time of its identifications (see CentrePSMs and RequirePeakAtScanTime). The fitted area is the reported quantity.

  2. Metabolic labelling

    For metabolically labeled samples the tool computes the m/z of the partner ion (the heavy form of a light identification, or the other way round) and quantifies that trace as well, so every row can carry a light and a heavy value.

Overview of the quantification process

InputsFiles and command-line flags

These inputs are needed to execute the tool:

Flag Meaning Comes from
-i one or more .mzlite files, or a directory MzMLToMzLite
-ii one or more .qpsm files, or a directory that is searched for *.qpsm PSMStatistics
-o the output directory, created when missing -
-d the SQLite peptide database PeptideDB
-p the parameter file in JSON QuantificationParams.json
-mf switch: pair files by base name instead of by position -
-dc switch: write diagnostic charts -
-z switch: pack the chart directory into a zip file and delete it -
-c number of runs quantified at the same time, default 1 -
  • The input must be .mzlite. A directory given to -i is also scanned for .mzML, but the tool opens every file with the mzlite reader and fails on anything else, so convert first. With one file per flag the pair is used as given. With lists, the n-th mzlite file is paired with the n-th qpsm file unless you pass -mf, which pairs files by their base name (run1.mzlite with run1.qpsm) and skips mzlite files without a partner.

  • For every run the tool writes run.quant, tab separated with a header and one row per quantified peptide ion.

  • Next to the identifiers and the best Q-value and PEP value of the ion it reports:

    • the fitted quantity (Quant_Light, Quant_Heavy)
    • the apex intensity
    • the fit parameters
    • the difference between identification and fitted scan time
    • and the extracted traces and the isotope pattern for both channels.
  • With -dc the tool writes a chart per peptide ion and the m/z correction chart mzErrorAndCorrection.html into <run>_plots. On successful archiving, -z writes <run>_plots.zip and removes that directory.

  • Delete an existing run.quant in the output directory before a rerun, the tool appends to it. Logs go to PSMBasedQuantification_log.txt and run_log.txt in the output directory.

Parameter fileDefault values and tool settings

General settings

Parameter Default Description
PerformLabeledQuantification Labeling.N15Labeling Unlabeled quantifies only the identified ion. N15Labeling quantifies the partner ion too and keeps rows where both fits pass the quality filter. N15LabelingOnly applies the filter to the heavy channel alone. Labelshift quantifies both and skips the filter.
XicExtraction.ScanTimeWindow 2.0 Half width of the scan time window around the identification that is extracted and searched for the peak, in the scan time unit of the run.
XicExtraction.MzWindow_Da Window.Estimate Half width of the m/z window of the trace, in Da, the same at every m/z. The trace is centred on the m/z corrected by an offset that the tool fits on the precursor m/z of the identifications, in ppm over retention time and m/z. Estimate uses four times the standard deviation of the difference between the measured and theoretical precursor m/z of the identifications, leaving out outliers. Window.Fixed 0.07 sets 0.07 Da.
XicExtraction.XicProcessing XicProcessing.Wavelet waveletParams Peak detection on the trace. The wavelet parameters are listed below. SecondDerivative is the other supported case. Gabor3D is in development and does not work currently.
XicExtraction.TopKPSMs None Some k keeps only the k identifications with the highest ModelScore per peptide ion when placing the scan time.
XicExtraction.IonMobilityWindow Window.Fixed 0.05 Not used, the tool does not read ion mobility.
BaseLineCorrection Some { MaxIterations = 10; Lambda = 6; P = 0.05 } Asymmetric least squares baseline subtraction on the trace before peak detection. None skips it.
QualityFilterLog2Limit 2.0 The quality filter keeps rows with -limit < log2 ((area / median area) / (apex / median apex)) < limit. Rows with a non-finite or non-positive area, apex or ratio are dropped.
CentrePSMs 3 Number of identifications per peptide ion, ranked by the summed intensity of their matched fragment ions, whose intensity weighted scan time centres the trace. It is applied after TopKPSMs, which ranks by ModelScore. At least 1.
RequirePeakAtScanTime true true quantifies only a peak that contains the weighted scan time of the identifications, so an ion without such a peak gets no row. The labeled partner needs a peak of its own trace that contains the fitted scan time of the ion. Without one it is not quantified, and the quality filter then drops the row with N15Labeling, and with N15LabelingOnly when the missing partner is heavy. false takes the peak with the closest apex when no peak contains the scan time.

The wavelet parameters:

Parameter Default Description
Borderpadding None Number of points padded at both ends of the trace. None derives it from the data.
BorderPadMethod Padding.BorderPaddingMethod.Random How the border points are filled (Random or Zero).
InternalPaddingMethod Padding.InternalPaddingMethod.LinearInterpolation How small gaps inside the trace are filled.
HugeGapPaddingMethod Padding.HugeGapPaddingMethod.Zero How gaps longer than HugeGapPaddingDistance are filled.
HugeGapPaddingDistance 100. Gap length from which a gap counts as huge.
MinPeakDistance Some 0.1 Peaks closer than this (in minutes) are merged. The distance is at least 1.5 times the median point spacing of the trace, which is also what None uses.
MinPeakLength Some 0.1 Smallest peak width the wavelet scales start at.
MaxPeakLength 1.5 Largest peak width the wavelet scales go up to.
NoiseQuantile 0.01 Quantile of the wavelet correlation used as noise level.
MinSNR 0.01 A peak needs a correlation above MinSNR times the noise level.

Default parameter file: Default JSON File

OutputsGenerated data and log files

QuantBasedAlignment, AlignmentBasedQuantification, AlignmentBasedQuantStatistics and JoinQuantPepIonsWithProteins read these .quant files.

The output table contains 47 columns. It uses the same layout as PSMBasedQuantificationTIMs. The columns are grouped by category below.

Identification and PSM confidence

Column Description
StringSequence Amino Acid sequence (one-letter code) from PeptideDB which matches the psm candidate
GlobalMod Indicator, if a peptide ion species is labeled or unlabeled
Charge Precursor ion charge state
PepSequenceID Unique identifier of the unmodified peptide sequence, which points to PeptideDB
ModSequenceID Unique identifier of the modified peptide sequence (including PTMs, e.g., methylation), which points to PeptideDB
PrecursorMZ Theoretical m/z of the ion plus the m/z correction, the m/z the XIC is centred on
MeasuredMass Mass that corresponds to PrecursorMZ, in Dalton
TheoMass Theoretical peptide mass in the spectrum (based on amino acid composition) in Dalton
AbsDeltaMass Absolute difference between MeasuredMass and TheoMass
MeanPercolatorScore Mean ModelScore of the identifications of the peptide ion
QValue Lowest q-value among the identifications of the ion in the .qpsm
PEPValue PEP of that identification
ProteinNames ProteinNames of that identification in the .qpsm

Light quantification (8)

Column Description
QuantMz_Light Mass-to-charge ratio from the extracted XIC of the unlabeled peptide ion
Quant_Light Peak area from the extracted XIC of the unlabeled peptide ion
MeasuredApex_Light Measured Apex of unlabeled peak
Seo_Light Standard Error of Prediction for quantification
Params_Light Describes the shape of the fitted peak for the unlabeled version of a peptide. Contains information about peak height, estimated peak time, and peak width.
Difference_SearchRT_FittedRT_Light Difference between the retention time originally determined using PSMs and the retention time calculated from the measured peak
KLDiv_Observed_Theoretical_Light Describes how well the measured unlabeled isotope pattern matches the theoretically expected pattern.
KLDiv_CorrectedObserved_Theoretical_Light Corrected Kullback–Leibler divergence of isotopic pattern of unlabeled peptide ion versus the measured pattern

Heavy quantification (8)

Column Description
QuantMz_Heavy Mass-to-charge ratio from the extracted XIC of the labeled peptide ion
Quant_Heavy Peak area from the extracted XIC of the labeled peptide ion
MeasuredApex_Heavy Measured Apex of labeled peak
Seo_Heavy Standard Error of Prediction of quantification
Params_Heavy Describes the shape of the fitted peak for the labeled version of a peptide. Contains information about peak height, estimated peak time, and peak width.
Difference_SearchRT_FittedRT_Heavy Difference between the retention time originally determined using PSMs and the retention time calculated from the measured peak
KLDiv_Observed_Theoretical_Heavy Describes how well the measured labeled isotope pattern matches the theoretically expected pattern.
KLDiv_CorrectedObserved_Theoretical_Heavy Corrected Kullback–Leibler divergence of isotopic pattern of labeled peptide ion versus the measured pattern

Light isotope patterns and traces (6)

Column Description
IsotopicPatternMz_Light M/z-distribution of isotope cluster in spectra for unlabeled peaks
IsotopicPatternIntensity_Observed_Light Observed intensity distribution of unlabeled isotopic clusters
IsotopicPatternIntensity_Corrected_Light Corrected intensity distribution of unlabeled isotopic clusters
RtTrace_Light Retention Time Profile of unlabeled peptide ion
IntensityTrace_Observed_Light Intensities at the monoisotopic m/z over the scan times in RtTrace_Light, drop-outs removed, before baseline subtraction
IntensityTrace_Corrected_Light The intensities of IntensityTrace_Observed_Light after ALS baseline subtraction, negative values set to 0

Heavy isotope patterns and traces (6)

Column Description
IsotopicPatternMz_Heavy M/z-distribution of isotope cluster in spectra for labeled peaks
IsotopicPatternIntensity_Observed_Heavy Observed intensity distribution of labeled isotopic clusters
IsotopicPatternIntensity_Corrected_Heavy Corrected intensity distribution of labeled isotopic clusters
RtTrace_Heavy Retention Time Profile of labeled peptide ion
IntensityTrace_Observed_Heavy Intensities at the monoisotopic m/z of the labeled partner over the scan times in RtTrace_Heavy, drop-outs removed, before baseline subtraction
IntensityTrace_Corrected_Heavy The intensities of IntensityTrace_Observed_Heavy after ALS baseline subtraction, negative values set to 0

Light / Heavy comparison and quantification source (2)

Column Description
Correlation_Light_Heavy Correlation calculated based on Pearson between unlabeled and labeled peaks
QuantificationSource If spectra are found over Alignments or Peptide Spectrum Matching

Alignment quality (2)

Alignment scores and associated q-values

Column Description
AlignmentScore NaN for PSM-based rows. For transferred rows, the probability from AlignmentBasedQuantStatistics that the quantified peak belongs to the ion.
AlignmentQValue 0 for PSM-based rows. For transferred rows, the q-value from AlignmentBasedQuantStatistics.

Ion Mobility

Column Description
IonMobility Ion mobility of the identifications, weighted by their matched fragment intensity, NaN without ion mobility
IonMobilityCentre Ion mobility the XIC of the row is summed around, NaN without an ion mobility window
Parameter generationJSON and parameter-generation examples

From the ProteomIQon repository root, build the library with dotnet build src/ProteomIQon/ProteomIQon.fsproj -c Release. Save the following as generate-params.fsx in that same directory and run dotnet fsi generate-params.fsx. The script writes the JSON file to the system temporary directory.

#r "nuget: FSharp.Stats, 0.6.0"
#r "nuget: Newtonsoft.Json, 13.0.4"
#r "./src/ProteomIQon/bin/Release/net10.0/ProteomIQon.dll"

open ProteomIQon
open ProteomIQon.Domain
open FSharp.Stats.Signal

let waveletParams : WaveletParameters =
    {
        Borderpadding          = None
        BorderPadMethod        = Padding.BorderPaddingMethod.Random
        InternalPaddingMethod  = Padding.InternalPaddingMethod.LinearInterpolation
        HugeGapPaddingMethod   = Padding.HugeGapPaddingMethod.Zero
        HugeGapPaddingDistance = 100.
        MinPeakDistance        = Some 0.1
        MinPeakLength          = Some 0.1
        MaxPeakLength          = 1.5
        NoiseQuantile          = 0.01
        MinSNR                 = 0.01
    }

let quantificationParams : Dto.QuantificationParams =
    {
        PerformLabeledQuantification = Labeling.N15Labeling
        XicExtraction =
            {
                ScanTimeWindow = 2.
                MzWindow_Da    = Window.Estimate
                XicProcessing  = XicProcessing.Wavelet waveletParams
                TopKPSMs       = None
                IonMobilityWindow = Window.Fixed 0.05
            }
        BaseLineCorrection = Some { MaxIterations = 10; Lambda = 6; P = 0.05 }
        QualityFilterLog2Limit = 2.
        CentrePSMs = 3
        RequirePeakAtScanTime = true
    }

// Replace the temp folder with your project folder.
let outputPath = System.IO.Path.Combine(System.IO.Path.GetTempPath(), "QuantificationParams.json")

Json.serializeAndWrite outputPath quantificationParams

Installing and executing the tool

  1. Installing

    Install the tool globally with the .NET CLI.

    dotnet tool install --global ProteomIQon.PSMBasedQuantification
  2. Execute

    proteomiqon-psmbasedquantification -i path/to/run.mzlite -ii path/to/run.qpsm -d path/to/database.db -o path/to/output -p path/to/QuantificationParams.json
  3. Running the tool for multiple runs, for three at the same time

    proteomiqon-psmbasedquantification -i path/to/run1.mzlite path/to/run2.mzlite path/to/run3.mzlite -ii path/to/run1.qpsm path/to/run2.qpsm path/to/run3.qpsm -d path/to/database.db -o path/to/output -p path/to/QuantificationParams.json -c 3
  4. Two directories, paired by file name

    proteomiqon-psmbasedquantification -i path/to/mzliteFolder -ii path/to/qpsmFolder -d path/to/database.db -o path/to/output -p path/to/QuantificationParams.json -c 3 -mf
  5. All Flags

    proteomiqon-psmbasedquantification --help